• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH TEST THROMBIN REAGENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH TEST THROMBIN REAGENT Back to Search Results
Model Number TEST THROMBIN REAGENT
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An outside of the united states (ous) customer contacted a siemens customer care center (ccc).Quality controls (qc) recovered in range for both lots of test thrombin reagent on the day of the event.Siemens evaluated qc using both lot 01552 and lot 02056 of test thrombin reagent.All measured qc values recovered within the assigned ranges, independent of the reagent lot used.Siemens also measured five fresh frozen plasma samples and five heparinized plasma samples using both lot 01552 and lot 02056 of test thrombin reagent.A low result bias was observed with lot 01552 compared to lot 02056; however, no product problem was identified.Lot 02056 demonstrates an improved sensitivity for heparin.
 
Event or Problem Description
The customer reported that higher thrombin time (tt) results were obtained on five patient samples on a sysmex cs-5100 system using test thrombin reagent lot 02056, compared to the recoveries from an alternate sysmex cs-5100 system using test thrombin reagent lot 01552.The customer indicated that the lower tt results were considered correct and were reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the different tt results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEST THROMBIN REAGENT
Common Device Name
TEST THROMBIN REAGENT
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
emil-von-behring- str. 76
marburg D-350 41
GM  D-35041
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
emil-von-behring- str. 76
marburg D-350 41
GM   D-35041
Manufacturer Contact
christopher aebig
511 benedict ave.
tarrytown, NY 10591
9144153450
MDR Report Key17996820
Report Number9610806-2023-00019
Device Sequence Number12429621
Product Code GJA
UDI-Device Identifier00842768012716
UDI-Public00842768012716
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/09/2025
Device Model NumberTEST THROMBIN REAGENT
Device Catalogue Number10446598
Device Lot Number02056
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/28/2023
Initial Report FDA Received Date10/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/10/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-