| Model Number |
VISUMAX |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
| Date of Event |
09/25/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The most possible root cause is a user error.There was too much liquid on the eye.As a consequence, the lenticule cut on the upper part is not sharp and not clearly defined which explains the difficult dissection.There is no indication that there was a malfunction of the zeiss device that contributed to the complications during lenticule removal.With the current information, there is no indication of any malfunction of the zeiss device which could have contributed to the reported surgical outcome.
|
| |
|
Event or Problem Description
|
|
It was reported by the surgeon that there was a complication with a lenticule removal, and sticky dissection.Lenticular interface was extremely sticky ; one quarter of side cut appeared to be missing.The doctor mentioned an injury: irregular astigmatism on patient cornea.
|
| |
|
Search Alerts/Recalls
|