• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 HI-WALL E1 LINER 36MM E; PROSTHESIS, HIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 HI-WALL E1 LINER 36MM E; PROSTHESIS, HIPS Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the surgeon operated strictly according to the standardized surgical technique, punched in the g7 52mm cup, used the corresponding acetabular liner trial mold, the trial mold was successfully implanted, and asked the off-stage nursing staff to open the corresponding high wall e1 acetabular liner 36mm size e.Surgeon strictly followed the e1 liner implantation surgical technique.The surgeon attempted to implant the liner two times into the cup and the liner would not insert.Later, it was replaced with the corresponding ceramic liner causing a 35-minute delay.
 
Additional Manufacturer Narrative
(b)(4).G2: foreign: china.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Updated: d2, d9, g3, g6, h2, h3, h6.Visual examination of the returned product identified the locking feature of the device has been indented and shows damage.No other damage is seen on the outer surface.It is unknown if the damage occurred prior to or during assembly attempts.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Review of complaint history found no additional related issues for this/these item(s) and the reported part and lot combination(s).Medical records were not provided.A definitive root cause cannot be determined.Complaint confirmed based on evaluation of the returned product.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL E1 LINER 36MM E
Common Device Name
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17997123
Report Number0001825034-2023-02483
Device Sequence Number5823189
Product Code PBI
UDI-Device Identifier00880304527102
UDI-Public(01)00880304527102(17)280115(10)7431309
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number010000935
Device Lot Number7431309
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2023
Supplement Date Received by Manufacturer01/05/2024
Initial Report FDA Received Date10/24/2023
Supplement Report FDA Received Date01/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
Patient Weight65 KG
-
-