Catalog Number 07398000160
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The analyzer serial number is (b)(6).The field service engineer (fse) ran a 21 cup precision test and mechanical and performance checks successfully.The investigation is ongoing.
 
Event or Problem Description
There was an allegation of questionable elecsys troponin t hs assay results for 1 patient sample on a cobas e 411 immunoassay analyzer.The initial troponin result was 14.87 ng/l.Another sample was collected from the patient that gave a result of 66.41 ng/l which prompted the customer to repeat the previous sample.The initial sample was repeated and the result was 66.92 ng/l.The repeat result was deemed correct.
 
Additional Manufacturer Narrative
Calibration was last performed on (b)(6) 2023.The calibration and qc recovery data provided was acceptable.The alarm trace did not contain a conspicuous event.No product problem was identified.The root cause of the event was found to be consistent with pre-analytical sample handling issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TROPONIN T STAT ASSAY
Common Device Name
IMMUNOASSAY METHOD, TROPONIN SUBUNIT
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17998273
Report Number1823260-2023-03416
Device Sequence Number12443765
Product Code MMI
UDI-Device Identifier04015630942695
UDI-Public04015630942695
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K051752
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07398000160
Device Lot Number726135
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/29/2023
Supplement Date Received by Manufacturer11/09/2023
Initial Report FDA Received Date10/24/2023
Supplement Report FDA Received Date11/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-