|
D10 concomitant products: 4.5pef, 4.5pef 4.5mm id paed med cuffless x1 (lot#unknown) 4.5pef, 4.5pef 4.5mm id paed med cuffless x1 (lot#unknown) 4.5pef, 4.5pef 4.5mm id paed med cuffless x1 (lot#unknown).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that all the components were assembled properly and the inner diameter of the tube was smeared with organic material which was stick to the surface.No black discoloration was observed.It was reported that there was a black discoloration inside the lumen of the tracheostomy tube.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the tube and accessories are supplied sterile and cannot be adequately resterilized.The tube and accessories are for single use and single patient use only.Attempts to resterilize the tube and accessories may result in product failure and increased risk to the patient.The tube and obturator are for single patient use.Duration should not exceed twenty-nine days.Covidien has not substantiated use of these devices beyond 29 days.Decisions about tracheostomy tube changes should be made by the responsible physician or designate using accepted medical techniques and judgment.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|