| Model Number |
MT22949 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the receiver display was frozen.Data was received but not investigated as data will not confirm the issue.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the receiver display was frozen.The product was evaluated.An external visual inspection was performed and passed.Receiver charge test was performed and passed.Receiver boot test was performed and passed.Functional test was performed and passed.A touchscreen manual button test was performed and passed.Internal visual inspection was performed and passed.The receiver log was downloaded and reviewed finding no error related to the complaint.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|