VERYAN MEDICAL LTD BIOMIMICS 3D VASCULAR STENT SYSTEM Back to Search Results
Catalog Number 142122-14
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Restenosis (4576)
Date of Event 12/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
This report is related to mdr number 3011632150-2023-00010.The patient was treated as part of the mimics 3d usa post-market observational study.On the (b)(6) 2021, the patient was implanted with one biomimics 3d (bm3d) stent (the study device), a 6.0 x 150mm stent which was used to treat a denovo lesion of the superficial femoral artery (sfa) distal third to proximal popliteal artery in the right leg.A contralateral approach was used and the lesion was prepared using pre-dilation with percutaneous transluminal angioplasty (pta) balloon and atherectomy.The treated segment was also post-dilated with pta.On the (b)(6) 2022, an event of occlusion (mdr 3011632150-2022-00105) was reported and it was treated with an additional non-study bm3d device (the subject of this report) which was placed in the segment between the sfa proximal third and sfa distal third.The stent details are unknown.The site reported that on (b)(6) 2022, a restenosis of treated segment (target lesion) was identified.It was reported as not related to the device and not related to the procedure but due to a worsening pre-existing condition.It was reported as target lesion related and required percutaneous intervention.The patient had a right leg extremity (rle) claudication rutherford class 3 (rc 3).A severe in-stent restenosis in the right sfa/popliteal.It was treated with laser atherectomy and pta/standard balloon angioplasty of the sfa proximal third to proximal popliteal artery on the (b)(6) 2022.It was reported as a target lesion/vessel revascularisation (tlr/tvr).This segment included areas of the vessel where both devices were implanted.The patient outcome was reported as resolved/recovered and the devices remain implanted.The details of this event were reviewed originally by veryan on (b)(6) 2023 where it was considered possibly related to the study device.The details of the non-study device were reviewed on (b)(6) 2023 and the event was considered possibly related to both devices.
 
Additional Manufacturer Narrative
This report is related to mdr number 3011632150-2023-00010.There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis leading to intervention and leg pain/claudication are listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
This report is related to mdr number 3011632150-2023-00010.There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis leading to intervention and leg pain/claudication are listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
 
Event or Problem Description
This report is related to mdr number 3011632150-2023-00010.The patient was treated as part of the mimics 3d usa post-market observational study.On the (b)(6) 2021, the patient was implanted with one biomimics 3d (bm3d) stent (the study device), a 6.0 x 150mm stent which was used to treat a denovo lesion of the superficial femoral artery (sfa) distal third to proximal popliteal artery in the right leg.A contralateral approach was used and the lesion was prepared using pre-dilation with percutaneous transluminal angioplasty (pta) balloon and atherectomy.The treated segment was also post-dilated with pta.On (b)(6) 2022, an event of occlusion (mdr 3011632150-2022-00105) was reported and it was treated with an additional non-study 6.0 x 125 mm bm3d device (the subject of this report) which was placed in the segment between the sfa proximal third and sfa distal third.The site reported that on (b)(6) 2022, a restenosis of treated segment (target lesion) was identified.It was reported as not related to the device and not related to the procedure but due to a worsening pre-existing condition.It was reported as target lesion-related and required percutaneous intervention.The patient had a right leg extremity (rle) claudication rutherford class 3 (rc 3).A severe in-stent restenosis in the right sfa/popliteal.It was treated with laser atherectomy and pta/standard balloon angioplasty of the sfa proximal third to the proximal popliteal artery on (b)(6) 2022.It was reported as a target lesion/vessel revascularisation (tlr/tvr).This segment included areas of the vessel where both devices were implanted.The patient outcome was reported as resolved/recovered and the devices remain implanted.The details of this event were reviewed originally by veryan on 09-jan-23 where it was considered possibly related to the study device.The details of the non-study device were reviewed on 25-sept-23 and the event was considered possibly related to both devices.Additional information received by the site on 27-feb-24 included the details of the non-study device that was implanted on (b)(6) 2022.
 
Additional Manufacturer Narrative
This report is related to mdr number 3011632150-2023-00010.There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis leading to intervention and leg pain/claudication are listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.Section a.1.Was updated to reflect the updated patient identifier, section b.5.Was updated with additional information, sections g.6.And h.2.Were updated with the type of report (follow-up 02) and the reason and section h.11.Was updated to reflect the changed sections in this report.
 
Event or Problem Description
This report is related to mdr number 3011632150-2023-00010.The patient was treated as part of the mimics 3d usa post-market observational study.On the (b)(6) 2021, the patient was implanted with one biomimics 3d (bm3d) stent (the study device), a 6.0 x 150mm stent which was used to treat a denovo lesion of the superficial femoral artery (sfa) distal third to proximal popliteal artery in the right leg.A contralateral approach was used and the lesion was prepared using pre-dilation with percutaneous transluminal angioplasty (pta) balloon and atherectomy.The treated segment was also post-dilated with pta.On 10-mar-22, an event of occlusion was reported and it was treated with an additional non-study 6.0 x 125 mm bm3d device (the subject of this report) which was placed in the segment between the sfa proximal third and sfa distal third.The site reported that on (b)(6) 2022, a restenosis of treated vessel (target vessel) was identified.It was reported as not related to the device and not related to the procedure but due to a worsening pre-existing condition.It was reported as not target lesion-related and required percutaneous intervention.The patient had a right leg extremity (rle) claudication rutherford class 3 (rc 3).A severe in-stent restenosis in the right sfa/popliteal.It was treated with laser atherectomy and pta/standard balloon angioplasty of the sfa proximal third to the proximal popliteal artery on (b)(6) 2022.It was reported as a target lesion/vessel revascularisation (tlr/tvr).This segment included areas of the vessel where both devices were implanted.The patient outcome was reported as resolved/recovered and the devices remain implanted.The details of this event were reviewed originally by veryan on 09-jan-23 where it was considered possibly related to the study device.The details of the non-study device were reviewed on 25-sept-23 and the event was considered possibly related to both devices.Additional information reviewed on 15-apr-24 included the clinical events committee (cec) determination that this event of restenosis was not related to the study device implanted.This event was the second event that had occurred in the study stent and required a tlr intervention and as a result, this event was determined to be due to the progression of the patient's underlying disease.The first event was an occlusion event that was reported in mdr 3011632150-2022-00105.The restenosis involved the non-study stented segment and this was the first restenosis of this stent so it was still considered possibly related to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Common Device Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Manufacturer (Section D)
VERYAN MEDICAL LTD
block 5
parkmore east business park
galway, H91 V 0TX
EI  H91 V0TX
Manufacturer Contact
alan mcdonagh
block 5
parkmore east business park
galway, H91 V-0TX
EI   H91 V0TX
MDR Report Key18001364
Report Number3011632150-2023-00129
Device Sequence Number15373004
Product Code NIP
UDI-Device Identifier05391526850480
UDI-Public(01)05391526850480(17)220702(11)201201(10)0000052898
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P180003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/02/2022
Device Catalogue Number142122-14
Device Lot Number0000052898
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2023
Supplement Date Received by Manufacturer02/27/2024
04/15/2024
Initial Report FDA Received Date10/25/2023
Supplement Report FDA Received Date03/26/2024
05/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASPIRIN; CLOPIDOGREL (PLAVIX)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexMale
Patient Weight234 KG
Patient EthnicityHispanic
-
-