| Model Number |
NEEDLE, TUOHY WITH OBTURATOR |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that, during an autocart procedure, metal debris was found in the harvested cartilage mass.There was no harm for patient, operator or third party reported.The surgery was finished successfully with the same device.It was not necessary to switch the surgical technique or do a second surgery.09-oct-2023 update dw: further information was provided that after the cartilage is cut with the sabre ar-7300sr and collected with the graftnet it was noticed that no metal particles were found in the cartilage mass at this moment.The metal abrasion only became visible in the joint when the cartilage mass was applied, indicating that the debris was caused by the syringe.The surgeon decided to use the mass with the particles to finish the surgery successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information; g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause can be attributed to user error due to excessive pressure used when drawing back the syringe and/or excessive friction during use.
|
| |
|
Search Alerts/Recalls
|