ARTHREX, INC. NEEDLE, TUOHY WITH OBTURATOR; ARTHROSCOPIC ACCESSORIES Back to Search Results
Model Number NEEDLE, TUOHY WITH OBTURATOR
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that, during an autocart procedure, metal debris was found in the harvested cartilage mass.There was no harm for patient, operator or third party reported.The surgery was finished successfully with the same device.It was not necessary to switch the surgical technique or do a second surgery.09-oct-2023 update dw: further information was provided that after the cartilage is cut with the sabre ar-7300sr and collected with the graftnet it was noticed that no metal particles were found in the cartilage mass at this moment.The metal abrasion only became visible in the joint when the cartilage mass was applied, indicating that the debris was caused by the syringe.The surgeon decided to use the mass with the particles to finish the surgery successfully.
 
Additional Manufacturer Narrative
Additional information; g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause can be attributed to user error due to excessive pressure used when drawing back the syringe and/or excessive friction during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEEDLE, TUOHY WITH OBTURATOR
Common Device Name
ARTHROSCOPIC ACCESSORIES
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18001813
Report Number1220246-2023-08348
Device Sequence Number15116075
Product Code NBH
UDI-Device Identifier00888867001817
UDI-Public00888867001817
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 03/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNEEDLE, TUOHY WITH OBTURATOR
Device Catalogue NumberABS-1001
Device Lot Number3061129650
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2023
Supplement Date Received by Manufacturer03/05/2025
Initial Report FDA Received Date10/25/2023
Supplement Report FDA Received Date03/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-