| Model Number |
37370DL |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The previous reportability decision for this event was reversed based on a retrospective review.The affected device has been requested for investigation by the manufacturer.Device was returned for investigation.Evaluation findings: the insulation is missing from the distal end hook.The hook is bent.The remaining insulation passed the hi-pot test.There are minor dents/scratches on the insulation.The seal is cracked at the proximal end.There is residue on the connector pin.There is temperature damage inside the distal end of the shaft.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic cholecystectomy the device 37370dl had broken inside the patient's body.The user had been able to retrieve all pieces that they could find, and no harm occurred to the patient.
|
| |
|