Model Number 37370DL
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The previous reportability decision for this event was reversed based on a retrospective review.The affected device has been requested for investigation by the manufacturer.Device was returned for investigation.Evaluation findings: the insulation is missing from the distal end hook.The hook is bent.The remaining insulation passed the hi-pot test.There are minor dents/scratches on the insulation.The seal is cracked at the proximal end.There is residue on the connector pin.There is temperature damage inside the distal end of the shaft.The event is filed under internal karl storz complaint id: (b)(4).
 
Event or Problem Description
It was reported that during a laparoscopic cholecystectomy the device 37370dl had broken inside the patient's body.The user had been able to retrieve all pieces that they could find, and no harm occurred to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGULATING AND DISSECTING ELECTRODE
Common Device Name
COAGULATING AND DISSECTING ELECTRODE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
christiane klaiber
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
MDR Report Key18004650
Report Number9610617-2023-00323
Device Sequence Number20742657
Product Code KNF
UDI-Device Identifier04048551155653
UDI-Public4048551155653
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K944862
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number37370DL
Device Catalogue Number37370DL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2023
Initial Report FDA Received Date10/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-