• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG MACRO NEEDLE HOLDER, 5 MM, 33 CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG MACRO NEEDLE HOLDER, 5 MM, 33 CM Back to Search Results
Model Number 26173KPF
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the needle holder 26173kpf broke in the procedure.No further information was provided about the procedure, but it was stated that the patient was unharmed.
 
Additional Manufacturer Narrative
The previous reportability decision for this event was reversed based on a retrospective review.The affected device has been requested for investigation by the manufacturer.Device was returned for investigation.Evaluation findings: broken jaws, residue, corrosion, resistance, misaligned ratchet, cracked epoxy.Evidence of use: a piece of the carbide coating is broken off the distal jaws of the needle holder.The broken piece was not returned with the instrument.There is residue and corrosion in the distal jaws.The release button has some resistance when depressing it.The ratchet lock in the handle is misaligned.The shaft connection to the handles has cracked epoxy.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MACRO NEEDLE HOLDER, 5 MM, 33 CM
Common Device Name
MACRO NEEDLE HOLDER, 5 MM, 33 CM
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
christiane klaiber
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
MDR Report Key18004656
Report Number9610617-2023-00322
Device Sequence Number15363051
Product Code GCJ
UDI-Device Identifier04048551098219
UDI-Public4048551098219
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K951625
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26173KPF
Device Catalogue Number26173KPF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2023
Initial Report FDA Received Date10/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-