• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. HD SCORPION NEEDLE (5 PACK); MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. HD SCORPION NEEDLE (5 PACK); MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number HD SCORPION NEEDLE (5 PACK)
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/29/2023
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2023, it was reported by a sales representative via sems-06045007 that an ar-13999hdn hd scorpion needle with megaloader, ve5, had an issue.The surgeon was completing an aflex 301 patch over a rotator cuff, side-side repair.The surgeon placed and tied 3 suturetapes, starting anterolateral and moving posteriorly.On attempting to place a final suture tape on the posterolateral corner of the patch/cuff tissue, the distal 8-9mm of the hd scorpion needle broke off, cutting the suture.On pulling the scorpion out of the joint, the broken needle tip flew out and landed on the back scrub table.The surgeon stated after the case, ¿i felt the needle bend as i fired it through the patch.¿ the case was completed using another ar-13999hdn hd scorpion needle and several packs of ar-7501 with a delay of 4-5 minutes while retrieving and opening a new scorpion hd needle.There was no additional anesthesia needed.This was discovered during rotator cuff repair with a patch augmentation procedure on (b)(6) 2023.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Complaint allegation is confirmed.One unpackaged ar-13999hdn was received for investigation.Functional testing was not conducted due to the damage to the instrument's tip.Visual evaluation found that the scorpion needled strip was bend and the tip was broken off.The most likely cause(s) for the reported failure can be a user error due to excessive force during use, misaligned insertion, and/or prying/levering the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD SCORPION NEEDLE (5 PACK)
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18005508
Report Number1220246-2023-08379
Device Sequence Number12422757
Product Code LXH
UDI-Device Identifier00888867374003
UDI-Public00888867374003
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHD SCORPION NEEDLE (5 PACK)
Device Catalogue NumberAR-13999HDN
Device Lot Number15108172
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2023
Supplement Date Received by Manufacturer12/20/2024
Initial Report FDA Received Date10/25/2023
Supplement Report FDA Received Date01/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-