| Model Number |
HD SCORPION NEEDLE (5 PACK) |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2023, it was reported by a sales representative via sems-06045007 that an ar-13999hdn hd scorpion needle with megaloader, ve5, had an issue.The surgeon was completing an aflex 301 patch over a rotator cuff, side-side repair.The surgeon placed and tied 3 suturetapes, starting anterolateral and moving posteriorly.On attempting to place a final suture tape on the posterolateral corner of the patch/cuff tissue, the distal 8-9mm of the hd scorpion needle broke off, cutting the suture.On pulling the scorpion out of the joint, the broken needle tip flew out and landed on the back scrub table.The surgeon stated after the case, ¿i felt the needle bend as i fired it through the patch.¿ the case was completed using another ar-13999hdn hd scorpion needle and several packs of ar-7501 with a delay of 4-5 minutes while retrieving and opening a new scorpion hd needle.There was no additional anesthesia needed.This was discovered during rotator cuff repair with a patch augmentation procedure on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.One unpackaged ar-13999hdn was received for investigation.Functional testing was not conducted due to the damage to the instrument's tip.Visual evaluation found that the scorpion needled strip was bend and the tip was broken off.The most likely cause(s) for the reported failure can be a user error due to excessive force during use, misaligned insertion, and/or prying/levering the device.
|
| |
|
Search Alerts/Recalls
|