CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 14FR; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS Back to Search Results
Model Number 8884715148
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/25/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.H3: device evaluation comment - the customer stated that the device will not be returned for evaluation because it was discarded.
 
Event or Problem Description
Customer reports: the patient has a kangaroo gastrostomy tube with y port.As per manufacturer's instructions, the g-tube balloon is deflated and re-inflated.However, the daily deflating and re-inflating, weakened the retention balloon and the patient experienced a rupturing of the retention balloon while the tube was in use.Staff noticed this after some time as the gastric tube was sliding out.Also, there was some friction related redness at the insertion site as the tube was able to slide in and out during use.It is unclear why the manufacturer's instruction's require daily vs weekly deflating and reflating of the gastric tube balloon.Unfortunately, the faulty g-tube was thrown out by staff during the exchange process, thus a lot number cannot be provided.The tube was also initially inserted in community (ltc).This is also not the first time that we have experienced this at the bedside.Unfortunately, with previous g-tube balloon rupture, staff did not report to this.There was redness at the insertion site indicative of friction irritation due to retention balloon failure.
 
Additional Manufacturer Narrative
The device history record review could not be performed since no lot number was provided from customer complaint report, and no photo sample/actual sample with a lot number was received.The physical sample was not returned; however, two images were submitted for reference.After performing a visual inspection, the sample analysis was performed on the photograph.Based on this analysis, the condition reported by customer was not observed in the images because it is necessary to see how the product works in order to make a determination, therefore the condition reported could not be confirmed.A gemba walk was done at the manufacturing facility focusing on the manufacturing equipment and process; a current process and controls were found to be followed correctly, including subassemblies, finished product assembly, packaging, and product inspections; however, there have been similar cases reported in the past, of which a formal corrective/preventative action (capa) was issued to the supplier to address the root cause and action plan.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GSTRO FEED TBE W/Y PRT 14FR
Common Device Name
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS
Manufacturer (Section D)
CARDINAL HEALTH
777 west street, mansfield, ma
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key18005801
Report Number9612030-2023-03843
Device Sequence Number17776674
Product Code PIF
UDI-Device Identifier10884521007222
UDI-Public10884521007222
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884715148
Device Catalogue Number8884715148
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2023
Supplement Date Received by Manufacturer10/24/2023
Initial Report FDA Received Date10/25/2023
Supplement Report FDA Received Date11/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-