| Catalog Number |
CDC-42703-P1A |
| Medical Device Problem Code |
Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported the guidewire was severely bent upon opening kit.No patient involvement with device was reported.
|
| |
|
Event or Problem Description
|
|
It was reported the guidewire was severely bent upon opening kit.No patient involvement with device was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer provided one photo for analysis.The complaint of a kinked guide wire was able to be confirmed by the photo.A complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.A device history record review was performed, and no relevant findings were identified.The customer report of a kinked guide wire was confirmed by visual inspection of the customer supplied photo.However, full complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|