• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. MIC-KEY GASTRIC-JEJUNAL FEEDING TUBE KIT (ENFIT); DH EF BALLOON TUBES PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. MIC-KEY GASTRIC-JEJUNAL FEEDING TUBE KIT (ENFIT); DH EF BALLOON TUBES PRODUCTS Back to Search Results
Model Number 8270-16-3.5-45
Medical Device Problem Codes Leak/Splash (1354); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history record is in-progress.The product involved in the report has been returned and the investigation remains in progress at this time.Initial assessment of sample return performed on 03oct2023 identified this event to be reportable.All information reasonably known as of 23 oct 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint comp(b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Event or Problem Description
It was initially reported on (b)(6) 2023, the tube has split below the balloon."when completing a tube change in radiology, loss of contrast noted and leakage back towards the gut.On removal, large split noted on the tube." based on sample evaluation initiated on 03oct2023 (regulatory alert date), the evaluation confirms that the damage to the tubing allowed cross-communication between the gastric and jejunal lumens.
 
Additional Manufacturer Narrative
The device history record for lot 30156416 was reviewed and the product was produced according to product specifications.The actual complaint product was returned for evaluation.The sample provided was evaluated, confirming a tear located in the fourth gastric skive.The tear includes damage to the inner septum, which separates the gastric and jejunal feed lumens.However, per dhr review and process assessment conducted, there were no activities or tools identified that could cause this type of damage in the tube component.Therefore, this seems to be a user-related problem since, as per the ifu, do not use excessive force to flush the tube.Excessive force can perforate the tube.Per vygon, the device was in use for an extended period of time (100 days); which indicates that the device did not show this defect at time of placement and the tube performed as intended.The incident was confirmed as reported.The root cause is inappropriate use.All information reasonably known as of 31 oct 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIC-KEY GASTRIC-JEJUNAL FEEDING TUBE KIT (ENFIT)
Common Device Name
DH EF BALLOON TUBES PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key18007872
Report Number9611594-2023-00150
Device Sequence Number15116119
Product Code KNT
UDI-Device Identifier00350770441016
UDI-Public00350770441016
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K922667
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/22/2023
Device Model Number8270-16-3.5-45
Device Catalogue NumberN/A
Device Lot Number30156416
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2023
Supplement Date Received by Manufacturer10/26/2023
Initial Report FDA Received Date10/25/2023
Supplement Report FDA Received Date11/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age17 YR
Patient SexFemale
-
-