| Catalog Number |
68LP-350 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Invagination (1336)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/03/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional additionally reported right side "fold failure.¿ the device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6b, h6.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: observed opening assessed as fold flaw opening.¿ crease/folding of implant: observed creases on the device.Additional observations: ¿ no other observation observed.No further actions are required since no manufacturing issue is observed.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Healthcare professional additionally reported right side "fold failure.¿ the device has been explanted.
|
| |
|