Catalog Number 5531G109
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Joint Laxity (4526)
Date of Event 10/03/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.The following devices were also listed in this report: cat# 5510f102: triathlon cr fem comp #1 r-cem: lot# unknown.Cat# 5550-g-278; triathlon symmetric x3 patella; lot# unknown.Cat# 5521-b-100; tri ts baseplate size 1; lot# unknown.Cat# 6197-9-001; simplex p with tobramycin 1 pack; lot# unknown.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Event or Problem Description
Patient reported knee "collapses" and would like to know if parts are subject to a recall.
 
Additional Manufacturer Narrative
Reported event: an event regarding instability involving a triathlon insert was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.-clinician review: a review of medical records with a clinical consultant indicated this product inquiry concerns a patient who underwent a cemented total knee arthroplasty and is now complaining that the knee "collapses." the only way i can confirm that the patient had the knee replacement was by the documentation in the above implant record.It is unclear what this event actually is.The only symptom described is that the knee is "collapsing." this could describe a knee that is unstable or a knee that has a loose body within it or some other mechanical or soft tissue problem.Without more information the root cause of this issue cannot be determined.The root cause of instability is usually ligamentous of laxity which could be as a result of surgical technique factors, patient factors such as activity level in bmi and ligament integrity, and possibly implant factors.The implant itself could not be implicated as a causality unless there was any material defect in the implant.More information would be needed in this case.[.] -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of medical records with a clinical consultant indicated this product inquiry concerns a patient who underwent a cemented total knee arthroplasty and is now complaining that the knee "collapses." the only way i can confirm that the patient had the knee replacement was by the documentation in the above implant record.It is unclear what this event actually is.The only symptom described is that the knee is "collapsing." this could describe a knee that is unstable or a knee that has a loose body within it or some other mechanical or soft tissue problem.Without more information the root cause of this issue cannot be determined.The root cause of instability is usually ligamentous of laxity which could be as a result of surgical technique factors, patient factors such as activity level in bmi and ligament integrity, and possibly implant factors.The implant itself could not be implicated as a causality unless there was any material defect in the implant.More information would be needed in this case.[.] the patient would like to know if her implants are subject to a recall.A review indicated " no pfa's were found for the catalog number and lot combinations".No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Patient reported knee "collapses" and would like to know if parts are subject to a recall.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INS SIZE1 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick, NJ NA
EI   NA
2018315000
MDR Report Key18009584
Report Number0002249697-2023-01263
Device Sequence Number7537974
Product Code MBH
UDI-Device Identifier07613327045581
UDI-Public07613327045581
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Catalogue Number5531G109
Device Lot NumberLFR037
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/03/2023
Supplement Date Received by Manufacturer01/08/2024
Initial Report FDA Received Date10/26/2023
Supplement Report FDA Received Date02/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient SexFemale
Patient Weight63 KG
-
-