• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL SHERIDAN ET TUBE, SHER-I-BRONCH, RS, 37 FR; TUBE, TRACHEAL/BRONCHIAL, DIFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL SHERIDAN ET TUBE, SHER-I-BRONCH, RS, 37 FR; TUBE, TRACHEAL/BRONCHIAL, DIFF Back to Search Results
Catalog Number 5-16137
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the doctor found cuff leakage when doing clinical setting before using on patient.Then changed new one, no impact on patient".No patient involvement.
 
Additional Manufacturer Narrative
Qn#(b)(4).Other remarks: corrected data:.
 
Additional Manufacturer Narrative
(b)(4).The reported complaint of "leak - balloon cuff" could not be confirmed based upon the sample received.The customer returned one unit 5-16137 et tube, sher-i-bronch, rs, 37 fr for investigation.The et tube was returned with its stylet.The et tube was visually examined with and without magnification.Visual examination of the returned sample revealed that the et tube appears used as there was biological material present on the device.No other defects or anomalies were observed.A functional inspection was performed to attempt to inflate and deflate the balloon cuffs on the returned et tube.A syringe was filled with air and attached to the valve of the tracheal (white) pilot balloon.Upon injection of air, both the pilot balloon and cuff immediately inflated.The same process was used to test the bronchial (blue) inflation line.Once again, both the pilot balloon and cuff immediately inflated.The syringe was then removed, and the cuffs and pilot balloons were inspected for leaks.No leaks were detected.The syringe was then reattached in order to deflate the balloons.All pilot balloons and balloon cuffs were able to deflate.The et tube was submerged underwater and an attempt to inflate the balloon cuffs was made to detect any leaks.Upon injection of air, no leaks were detected.No functional issues were found with the returned sample.The ifu for this product states, "the cuff inflation system (cuffs, pilot balloons, valves) should be checked by inflation and deflation prior to use.If cuff is damaged, the tube should not be used.Care should be taken to avoid damaging the cuffs during intubation.If any one of the cuffs is damaged, the tube should not be used." "cuff pressure should be monitored routinely to assure that an adequate seal is maintained and that the cuffs have not become over-inflated." "deflate all cuffs prior to repositioning the tube.Movement of the tube with cuffs inflated could result in patient injury or in damage to the cuffs, requiring a tube change.Verify the position of the tube after each repositioning." a device history record review was performed, and no relevant findings were identified.The returned et tube was able to inflate and deflate with no difficulty.All et tubes are 100% inspected for both inflation and deflation at the time of manufacturing.No functional issues were found with the returned device.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that "the doctor found cuff leakage when doing clinical setting before using on patient.Then changed new one, no impact on patient".No patient involvement.
 
Additional Manufacturer Narrative
Qn# (b)(4).In section d provided the full udi #.Additionally added the brand name of the device.**udi related data quality updates only**.
 
Event or Problem Description
It was reported that "the doctor found cuff leakage when doing clinical setting before using on patient.Then changed new one, no impact on patient".No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHERIDAN ET TUBE, SHER-I-BRONCH, RS, 37 FR
Common Device Name
TUBE, TRACHEAL/BRONCHIAL, DIFF
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
marling heaton
3015 carrington mill blvd
morrisville 27560
MDR Report Key18011713
Report Number3003898360-2023-01447
Device Sequence Number7469269
Product Code CBI
UDI-Device Identifier04026704397662
UDI-Public(01)04026704397662(11)190925(17)240924(10)73J1900414(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K180253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/24/2024
Device Catalogue Number5-16137
Device Lot Number73J1900414
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2023
Supplement Date Received by Manufacturer11/28/2023
Not provided
Initial Report FDA Received Date10/26/2023
Supplement Report FDA Received Date11/29/2023
07/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/25/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.; N/A.; N/A.
Patient SexUnknown
-
-