MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER
|
Back to Search Results |
|
| Model Number |
PED2-500-12 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that despite proper manipulation, distal end of the pipeline would not open.The pipeline was pulled and deployed it on the back table.Even after deployment, a portion of the pipeline was stuck/hung up on itself (distal 1/3) not allowing it to expand to its nominal width.The physician played with the pipeline and inspected it which did force it out to its nominal 5.25mm width, but visible inspection will allow you to see lack of symmetry along the body of the pipeline.Additionally it appeared as though the wings would not separate from the distal portion of the pipeline.The pipeline was not positioned in a bend and more than 50% of the pipeline was deployed when it failed to open.The pipeline was resheathed less than or equal to two times.The pipeline was not used for an off-label use.The pipeline and any accessories were prepared as indicated in the instructions for use (ifu).No patient symptoms or further complications were reported as a result of this event.He patient was undergoing surgery for treatment of a saccular, unruptured left para ophthalmic aneurysm with a max diameter of 3.4mm and a 2.6mm neck diameter.The landing zone artery size measurements were 3.86mm proximally and 4.87mm distally.It was noted the patient's vessel tortuosity was moderate.Ancillary devices include a navien 058 guide catheter, phenom 27 microcatheter, and an aristotle 24 guidewire.
|
| |
|
Search Alerts/Recalls
|
|
|