Section a2, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section b3 - date of event: date unknown, as information was requested but not provided.Section d4 - expiration date: does not apply, re-usable device.Section d6a - implant date: not applicable.Phaco tip is not an implantable device.Section d6b - explant date: not applicable.Phaco tip is not an implantable device; therefore, not explanted.Section e1 - telephone number:(b)(6).Section h3 - other (81): the phaco tip was not returned for evaluation.Therefore, a failure analysis of the complaint device could not be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review and possible product return and evaluation, if there is any further relevant information a supplemental medwatch will be filed.Section h4 - device manufacture date: unknown, not provided.Attempts have been made to obtain missing information.However, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
|
Additional information: section d9 - device available for evaluation? yes.Section d9 - date returned to manufacturer: 07-dec-2023.Section h3 - device evaluated by manufacturer? yes.Device evaluation: the phaco tip was returned for evaluation.The mentioned lot returned with the cannula broken off at the base.Top view of the device shows that a part of cannula left inside the base is oval shape instead of circular shape and off center of the base, which is a sign of external force being applied to one side and breaking the cannula off.Therefore, functional testing was performed, and the device is performing within specifications.The reported issue could not be confirmed.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|