EXACTECH, INC. EQUINOXE CAGE GLENOID LARGE, BETA; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Model Number |
EQUINOXE CAGE GLENOID LARGE, BETA |
| Medical Device Problem Codes |
Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
10/03/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient, initial shoulder implanted on (b)(6) 2016, underwent a revision procedure on (b)(6) 2023, approximately 7 years 8 months post the initial procedure.The patient had presented with pain.The glenoid component was stable but was removed easily except the central cage remained in the bone.A trephine reamer and thread in extractor were used to remove the glenoid.Significant wear of the glenoid was visually confirmed prior to its removal.The stem was planned to be kept, but was found to be loose, so removal was performed and a conversion to a competitor¿s components was done with significant bone grafting on glenoid.There were no issues with the surgery.X-rays and device images provided.Explants are not available for return.Reason unknown.No further information.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitants: 3885932 300-01-13 - equinoxe, humeral stem primary, press fit 13mm.4100198 300-10-45 - equinoxe replicator plate 4.5mm o/s.4238199 300-20-02 - equinox square torque define screw drive kit.2832614 310-01-50 - equinoxe, humeral head short, 50mm (beta).4245253 321-20-00 - equinoxe reverse shoulder drill kit.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: b3, d1, h6.Mdr section codes updated/corrected: a, b, c, d, g.The reason for the revision cannot be confirmed but may have been due to an insufficient bond between the glenoid component and the bone and/or the humeral component and the bone leading to aseptic (non-infected) loosening of the glenoid and humeral components respectively, reported wear of the glenoid component, and/or the inclusion of the polyethylene in the packaging recall.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient, initial shoulder implanted, underwent a revision procedure, approximately seven years eight months post the initial procedure.The patient had presented with pain.The glenoid component was stable but was removed easily except the central cage remained in the bone.A trephine reamer and thread in extractor were used to remove the glenoid.Significant wear of the glenoid was visually confirmed prior to its removal.The stem was planned to be kept, but was found to be loose, so removal was performed and a conversion to a competitor¿s components was done with significant bone grafting on glenoid.There were no issues with the surgery.X-rays and device images provided.Explants are not available for return.Reason unknown.No further information.
|
| |
|