• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOPERSURGICAL, INC. TRANSWARMER MATTRESS; REACTIVE-GEL HEATING PAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOPERSURGICAL, INC. TRANSWARMER MATTRESS; REACTIVE-GEL HEATING PAD Back to Search Results
Model Number 20421
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Customer has indicated that the product was discarded and can not be returned to cooper surgical for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that while using the device, the patient suffered a burn of unknown degree.The patient condition is unknown as they were transferred to another facility.The customer stated that the device was used in conjunction with an additional heat source.Device was discarded by customer.No additional information is available.1216677-2023-00142 20421 transwarmer 2023-10-0000176.
 
Event or Problem Description
No additional information is available.
 
Additional Manufacturer Narrative
Distribution history the complaint product was purchased from pristech products inc.By csi.Manufacturing record review manufacturing record review not applicable to this product.Incoming inspection review a review of the incoming inspection record could not be performed because the complaint product lot/serial number was not provided.Should the complaint product lot/serial number be provided going forward, the incoming inspection report will be reviewed, and this complaint amended accordingly.Service history record service history not applicable for this product.Historical complaint review a review of the 2-year complaint history showed similar reported complaint conditions.1 complaint reported a burn to an infant, however, the complaint could not be confirmed.Product receipt the complaint product has not been returned to cooper surgical.Visual evaluation of the complaint product could not be completed as the complaint product has not been returned to cooper surgical.Functional evaluation of the complaint product could not be completed as the complaint product has not been returned to cooper surgical.Note: verified with the customer per correspondence that this device was used with an additional heat source, the omni giraffe bed despite the warning indicated on the device and ifu: do not apply an additional heat source to the infant when using this product.Verified both the product and the ifu state: do not leave on skin longer than 10 minutes.Root cause root cause not applicable as the complaint condition was not confirmed.Cooper surgical will continue to monitor this complaint condition for trends.No further corrective action is necessary, as the complaint condition was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSWARMER MATTRESS
Common Device Name
REACTIVE-GEL HEATING PAD
Manufacturer (Section D)
COOPERSURGICAL, INC.
95 corporate drive
trumbull CT 06611
Manufacturer (Section G)
COOPERSURGICAL, INC.
75 corporate drive
trumbull
CT 06611
Manufacturer Contact
michael marone
50 corporate drive
trumbull, CT 06611
4752651582
MDR Report Key18012821
Report Number1216677-2023-00142
Device Sequence Number7469560
Product Code IMD
UDI-Device Identifier00888937025767
UDI-Public(01)00888937025767
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K934631
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number20421
Device Catalogue Number20421
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/06/2023
Supplement Date Received by Manufacturer12/08/2023
Initial Report FDA Received Date10/26/2023
Supplement Report FDA Received Date12/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-