• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 1012536-80
Medical Device Problem Codes Break (1069); Malposition of Device (2616)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 10/04/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion right superficial femoral artery.Pre-dilatation was performed with an unspecified 5.0x100mm balloon catheter.An unspecified absolute pro self-expanding stent system (sess) was being deployed; however, the stent jumped forward 2cm proximal to treatment area.The stent was retracted distally, but the stent fractured and there was residual stenosis after an 8mm balloon angioplasty.An unspecified 8x39mm balloon-expandable stent was placed over the fractured stent and no residual stenosis was demonstrate after placement.There was no adverse patient sequela and no clinically significant delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.As there was no damage noted to the device during the inspection prior to use, it is possible that during deployment interaction with the anatomy and/or other devices resulted in a build-up of tension within the shaft lumens of the delivery system; thus resulting in a spring like release of the stent resulting in the reported malposition of device (stent jumped forward 2cm proximal to treatment area).Manipulation of the compromised device resulted in the reported stent break; however this cannot be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.The treatment(s) appears to be related to the operational context of the procedure as reportedly an unspecified 8x39mm balloon-expandable stent was placed over the fractured stent and no residual stenosis was demonstrate after placement.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18013229
Report Number2024168-2023-11889
Device Sequence Number7548640
Product Code NIP
UDI-Device Identifier08717648176166
UDI-Public08717648176166
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Catalogue Number1012536-80
Device Lot Number3052262
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2023
Supplement Date Received by Manufacturer02/19/2024
Initial Report FDA Received Date10/26/2023
Supplement Report FDA Received Date02/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-