| Model Number |
SHEATH,HF,TAP/FEN,2 STPCK FOR 4MM SCOPE |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed as the evaluation revealed that the stopcock is broken off at the weld.A most likely cause is high force application, improper use / handling or a drop of the product.Further the distal end is damaged and has a dent.
|
| |
|
Event or Problem Description
|
|
It was reported that the stopcock of the sheath is defective.There was no harm for patient, operator or third party reported.No further information received.
|
| |
|
Search Alerts/Recalls
|
|
|