ARTHREX, INC. SHEATH,HF,TAP/FEN,2 STPCK FOR 4MM SCOPE; ARTHROSCOPE AND ACCESSORIES Back to Search Results
Model Number SHEATH,HF,TAP/FEN,2 STPCK FOR 4MM SCOPE
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported event was confirmed as the evaluation revealed that the stopcock is broken off at the weld.A most likely cause is high force application, improper use / handling or a drop of the product.Further the distal end is damaged and has a dent.
 
Event or Problem Description
It was reported that the stopcock of the sheath is defective.There was no harm for patient, operator or third party reported.No further information received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHEATH,HF,TAP/FEN,2 STPCK FOR 4MM SCOPE
Common Device Name
ARTHROSCOPE AND ACCESSORIES
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18013323
Report Number1220246-2023-08396
Device Sequence Number15320821
Product Code NBH
UDI-Device Identifier00888867198579
UDI-Public00888867198579
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSHEATH,HF,TAP/FEN,2 STPCK FOR 4MM SCOPE
Device Catalogue NumberAR-3373-4002
Device Lot Number1691759
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2023
Initial Report FDA Received Date10/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/16/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-