| Model Number |
RFX072-115-08MP |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that during the interventional treatment of the aneurysm, the surgeon used a microcatheter to guide the intermediate catheter to be pushed to go upper.It was found that the intermediate catheter was kinked in the distal portion and the intermediate catheter could not be pushed to go upper.After it was withdrawn from the body, it was visible to the naked eye that the tube body had obvious fractures.No patient symptoms or further complications were reported as a result of this event.The device and any accessories were prepared as indicated in the ifu.The catheter was flushed as indicated in the ifu.The patient was undergoing surgery for interventional treatment of an aneurysm.The access vessel was the femoral artery with a diameter of 4.5mm.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis #(b)(4): ¿ as found condition: the navien catheter was returned for analysis within a shipping box and a plastic bio-pouch.¿ damage location details: no damages were found with the navien catheter hub.The navien catheter body was found damaged (broken) at ~18.5cm from the catheter distal tip.The navien catheter distal tip was found in good condition.¿ testing/analysis: none ¿ conclusion: based on the device analysis and reported information, the customer¿s ¿difficult navigation¿, ¿catheter kink/damage¿, and ¿catheter separation/break¿ reports were confirmed.Damage to the navien catheter can occur if advanced against resistance during navigation.Difficult navigation can occur due to patient vessel tortuosity.However, the cause could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|