| Model Number |
105581US |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that during implant pump preparation, the operating room nurse observed outflow graft 8928557 from implant kit: (b)(6) had a visible fold at the end of the bend relief.The surgeon was concerned this would impinge on the function of the outflow graft so they used outflow graft 8408593 from implant kit: (b)(6) to replace outflow graft 8928557.It was observed that outflow graft 8408593 also had a very slight fold at the end of the bend relief, however it was still implanted with pump: (b)(6).There were no patient consequences.Related manufacturer reference number: (b)(4) - outflow graft 8928557.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s investigation conclusion: review of the submitted photos confirmed the reported event.A manufacturing assessment (ma) task was initiated to address deformation of the outflow graft bend relief observed during implant.The outer layer of the outflow graft bend relief is applied by abbott¿s supplier, and the evaluation performed by the supplier identified no root cause.It was suspected that that the issue was not detected during inspection/verification as required in the supplier's corresponding procedures.An ecar was issued to the supplier for notification of the observed issue.An update to the receiving inspection and manufacturing procedures was initiated to provide clarity to the written inspection criteria.Relevant sections of the device history records for the outflow graft, lot number 8408593, and bend relief, lot number 8257830 were reviewed and showed no deviations from the manufacturing and qa (quality assurance) specifications.There were no ncmrs (non-conformance material report), or re-works related to the reported event.The heartmate 3 lvas instructions for use (ifu), rev.C is currently available.Section 5 of this ifu contains instructions for unpacking and preparing the sealed outflow graft.Section 5 also states that during the implant process, a complete backup system (implant kit and external components) must be available on-site and in close proximity for use in an emergency.No further information was provided.The manufacturer is closing the file on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d1 brand name: corrected.Section d4 catalog number: corrected.Section d4 primary udi number: corrected.No further information was provided.The manufacturer is closing the file on this event.
|
| |
|
Search Alerts/Recalls
|