MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number CAR-505
Medical Device Problem Codes Material Separation (1562); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned cartridge was evaluated and confirmed a break in the venous line luer.A device history record (dhr) review was conducted for the reported lot number and revealed the product was released for distribution having met quality and manufacturing specifications and requirements.All treatments must be administered under a physician''s prescription and performed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician so that alarms and harmful conditions can be responded to promptly.The instructions for use states "make sure mated luer-connectors are secure but do not over-tighten, especially when connections are wet.".
 
Event or Problem Description
A report was received on 05 oct 2023 from the health professional of a critical care patient of unspecified pathology, who stated the cartridge luer connector broke and blood could not be returned to the patient during a continuous veno-venous hemofiltration (cvvh) treatment on (b)(6) 2023.Additional information was received on 06 oct 2023 from the health professional who stated there was no report of adverse impact to the patient or report of medical intervention being required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Common Device Name
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key18014341
Report Number3003464075-2023-00094
Device Sequence Number15340618
Product Code KDI
UDI-Device IdentifierM535CAR5050
UDI-Public+M535CAR5050/$$082430478004Z
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K133547
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2024
Device Model NumberCAR-505
Device Catalogue NumberCAR-505
Device Lot Number30478004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-