| Model Number |
CAR-505 |
| Medical Device Problem Codes |
Material Separation (1562); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned cartridge was evaluated and confirmed a break in the venous line luer.A device history record (dhr) review was conducted for the reported lot number and revealed the product was released for distribution having met quality and manufacturing specifications and requirements.All treatments must be administered under a physician''s prescription and performed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician so that alarms and harmful conditions can be responded to promptly.The instructions for use states "make sure mated luer-connectors are secure but do not over-tighten, especially when connections are wet.".
|
| |
|
Event or Problem Description
|
|
A report was received on 05 oct 2023 from the health professional of a critical care patient of unspecified pathology, who stated the cartridge luer connector broke and blood could not be returned to the patient during a continuous veno-venous hemofiltration (cvvh) treatment on (b)(6) 2023.Additional information was received on 06 oct 2023 from the health professional who stated there was no report of adverse impact to the patient or report of medical intervention being required.
|
| |
|