• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEVRO CORP. SENZA; NEVRO SENZA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEVRO CORP. SENZA; NEVRO SENZA Back to Search Results
Model Number LEAD1058-50B
Medical Device Problem Codes Material Frayed (1262); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Nevro is awaiting the return of the device.The manufacturing records were reviewed and no relevant nonconformities were found.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Event or Problem Description
It was reported that the patient had a frayed lead insulation near the ipg connection end of the lead.An hcp assessment regarding the cause of the reported event was not available.The device was replaced and there have been no reports of further complications regarding this event.
 
Additional Manufacturer Narrative
The device was returned and analyzed. the investigation found that the returned lead had damage along the lead body with exposed conductor material from a fractured cable protruding out of the body.The patient device history was limited, no device history information after about may 2021.Because the device statistics data showed that all contact impedances were within specification at the time of implant, we can conclude that the lead was most likely manufactured per specification.Without more device history from the patient¿s ipg it cannot be determined when the cable fractured.There is not enough evidence to determine if the lead had a protruding frayed cable while inside of the patient.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENZA
Common Device Name
NEVRO SENZA
Manufacturer (Section D)
NEVRO CORP.
1800 bridge parkway
redwood city CA 94065
Manufacturer Contact
zulfiqar ali
1800 bridge parkway
redwood city, CA 94065
6505317627
MDR Report Key18014580
Report Number3008514029-2023-00649
Device Sequence Number12658051
Product Code LGW
UDI-Device Identifier00813426020022
UDI-Public00813426020022
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P130022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/02/2019
Device Model NumberLEAD1058-50B
Device Lot Number94386222
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2023
Supplement Date Received by Manufacturer01/31/2024
Initial Report FDA Received Date10/26/2023
Supplement Report FDA Received Date02/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/02/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-