| Catalog Number |
252303103 |
| Medical Device Problem Codes |
Corroded (1131); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the femoral trial has rusted as the hospital used the wrong detergent.Patient status/ outcome / consequences , was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study --> unknown, (b)(4).Device property of none, device in possession of none, (b)(4).Device property of none, device in possession of none, (b)(4).Device property of none, device in possession of none, (b)(4).Device property of none, device in possession of none by checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary : items rusted due to wrong detergent used at hospital.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure, or patient details available the product was not returned to depuy synthes, however photos were provided for review.See attachment [20231023_165555.Jpg, 20231023_165539.Jpg].The photo investigation revealed that the attune rev t augtrl has signs of severe corrosion, the condition can be attributed to the improper maintenance, as was mentioned at the event description.Refer to ifu-0902-00-836 for cleaning and sterilization process.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the attune rev t augtrl would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.¿ as part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.¿ device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: items rusted due to wrong detergent used at hospital.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure, or patient details available.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that attune rev t augtrl 15mm sz3lm presents signs of corrosion all over the surface, no other defects that could have contributed to the reported complaint was observed.Based on the available information, the observed condition of the device can be attributed to an improper maintenance.Refer the ifu-0902-00-836 for cleaning and sterilization process.The overall complaint was confirmed as the observed condition of the attune rev t augtrl 15mm sz3lm would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to maintenance, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: items rusted due to wrong detergent used at hospital.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure, or patient details available the product was not returned to depuy synthes, however photos were provided for review.See attachment [(b)(4)].The photo investigation revealed that the attune rev t augtrl has signs of severe corrosion, the condition can be attributed to the improper maintenance, as was mentioned at the event description.Refer to ifu-0902-00-836 for cleaning and sterilization process.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the attune rev t augtrl would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|
|