| Model Number |
ORT300 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An imris serivce engineer assessed the operating room table at the customer site the same day (b)(6) 2023 and determined the acutator component within the table's rotational locking mechanism caused the malfunction.The service engineer removed and replaced the actuator component and functionally tested the table after the repair.The actuator was returned to the manufacturer and an internal corrective action has been initiated to further address this issue.
|
| |
|
Event or Problem Description
|
|
The customer reported they were not able to lock the rotational position in the ort300 operating room table.This occurred during equipment cleaning and had no procedural or patient involvment.
|
| |
|