| Catalog Number |
SSLP-300 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported a right side rupture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, broken device assessed as unidentified (tear) opening (shell thickness within specification) and missing shell assessed as inconclusive.No additional observations are performed.No further actions are required since no issue in the manufacturing of the device is observed.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side rupture.Device has been explanted and replaced.
|
| |
|