Section b3: date of event: the exact date is unknown, not provided.The patient indicated that after the lens was implanted her symptoms started; therefore, the best estimate date is on or after (b)(6) 2022.Section d6b: if explanted; give date: n/a (not applicable).The lens remains implanted.Device evaluation: the product testing could not be performed as the product was not returned for evaluation.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed no other complaints were received for this po.Conclusion: based on the investigation, no malfunction or product deficiency was identified.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
|
|
A female patient who had bilateral lens implantation reported of halos, starbursts, and double vision on both eyes which thinks are due to mechanical issues with the lenses.The patient stated that it is hard to work, drive and read, that she is very disappointed.The patient has astigmatism but no other health conditions.The patient reported the symptoms to her doctor and was told to wait a year to get used to the lenses; however, she still has the symptoms.The patient does work, read, and drive, but stated that it is difficult.The patient does limited driving at nighttime from work to home which takes the side streets as she thinks it is dangerous for her to drive more to other places at nighttime.No surgical or medical interventions have been performed and both lenses remain implanted.The patient has seen a new doctor who has offered to explant the lenses and the patient is scheduled for the right eye lens explant on (b)(6) 2023, and for the left eye in (b)(6) 2023.No further information was provided.This report is for the patient¿s right eye.A separate report is being submitted for the patient¿s left eye.
|