• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW AGBA PI MIDLINE 2L: 5.5FR X 15CM; MIDLINE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW AGBA PI MIDLINE 2L: 5.5FR X 15CM; MIDLINE CATHETER Back to Search Results
Catalog Number PR-41552-BAS
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that: when attempting to separate the sheath using the plastic tabs at the top of the sheath the two halves did not separate.One of the plastic tabs broke and a sterile forceps was needed to separate the sheath.This was successful and the procedure was completed.There was no patient injury or consequence.
 
Event or Problem Description
It was reported that: when attempting to separate the sheath using the plastic tabs at the top of the sheath the two halves did not separate.One of the plastic tabs broke and a sterile forceps was needed to separate the sheath.This was successful and the procedure was completed.There was no patient injury or consequence.
 
Additional Manufacturer Narrative
Qn# (b)(4).The report that a peel-away tabs broke off was confirmed through complaint investigation.Visual analysis revealed that the sheath was separated directly adjacent to the hub.Despite the damage, the sheath met all relevant dimensional requirements.Based on the customer report, the sample received, and the comments from r & d/mfg, manufacturing caused or contributed to this event.A device history record review was performed, and relevant findings were identified.For eb-01052-002 with batch 34c22c0032, investigations were created regarding peel away sheath extrusion defects.An investigation was initiated to further investigate the issue.The investigation is still on-going.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW AGBA PI MIDLINE 2L: 5.5FR X 15CM
Common Device Name
MIDLINE CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
mariah mackinnon
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18020690
Report Number9680794-2023-00809
Device Sequence Number6585729
Product Code PND
UDI-Device Identifier10801902136586
UDI-Public10801902136586
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K161313
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/29/2024
Device Catalogue NumberPR-41552-BAS
Device Lot Number13F22K0731
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2023
Supplement Date Received by Manufacturer11/20/2023
Initial Report FDA Received Date10/27/2023
Supplement Report FDA Received Date11/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-