• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. TORNIER FLEX SHOULDER SYSTEM HEAD HI OFFSET 39MMX14MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. TORNIER FLEX SHOULDER SYSTEM HEAD HI OFFSET 39MMX14MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number DWF139
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 10/03/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Total shoulder arthroplasty was performed for a patient with osteoarthritis of right shoulder.The tendency of dislocation of the humeral nail was observed on one week post-op.Manual reduction was performed for work around on that day and revision surgery was performed on 2 days later.The doctor commented it was thought that dislocation occurred because the humeral stem was implanted in a state where the anteversion angle of it was wider than expected.
 
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.H3 other text : device not returned.
 
Event or Problem Description
Total shoulder arthroplasty was performed for a patient with osteoarthritis of right shoulder.The tendency of dislocation of the humeral nail was observed on one week post-op.Manual reduction was performed for workaround on that day and revision surgery was performed on 2 days later.The doctor commented it was thought that dislocation occurred because the humeral stem was implanted in a state where the anteversion angle of it was wider than expected.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A device inspection was not possible since the affected device was not returned for investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER FLEX SHOULDER SYSTEM HEAD HI OFFSET 39MMX14MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
WRIGHT MEDICAL CORK (TORNIER ORTHOPEDICS IRELAND LTD)
harnetts cross
macroom, co. cork NA
EI   NA
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18022260
Report Number3000931034-2023-00376
Device Sequence Number6538563
Product Code KWS
UDI-Device Identifier03700434021783
UDI-Public03700434021783
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K190521
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date07/27/2026
Device Catalogue NumberDWF139
Device Lot NumberAD7789
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/03/2023
Supplement Date Received by Manufacturer12/04/2023
Initial Report FDA Received Date10/27/2023
Supplement Report FDA Received Date12/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-