• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR DRAGONFLY¿ OPTIS¿ IMAGING CATHETER; DIAGNOSTIC INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR DRAGONFLY¿ OPTIS¿ IMAGING CATHETER; DIAGNOSTIC INTRAVASCULAR CATHETER Back to Search Results
Catalog Number C408646
Medical Device Problem Codes Break (1069); Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported the dragonfly optis imaging catheter, could not be inserted onto the guidewire.The imaging catheter was broken.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Visual analysis and additional tests were performed on the returned device.The reported insertion difficulty and break were unable to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.A query of the complaint handling database for the reported lot was performed and revealed there is no indication of a lot level quality product issue.The investigation was unable to determine a cause for the reported difficulty inserting (guidewire into catheter¿s guidewire exchange minirail) and reported break.The catheter was returned with contrast/blood within the guidewire exchange lumen, which was removed with water.After removing the blood/contrast, the guidewire was able to be inserted fully through the minirail and without difficulty.It may be possible that during prep the user inadvertently had dried contrast or blood within the exchange lumen¿which would obstruct the guidewire insertion until removed.It could also be that the guidewire employed was damaged or its condition may have contributed to the difficulty inserting into the catheter; however, this could not be confirmed.There was an optical fiber break noted during investigation, which could not be directly attributed to the reported event.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.E1 phone number updated from (b)(6) due to 3 digit country code update.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY¿ OPTIS¿ IMAGING CATHETER
Common Device Name
DIAGNOSTIC INTRAVASCULAR CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT MEDICAL REG#3009600098
4 robbins rd
westford MA 01886
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18024801
Report Number2024168-2023-11943
Device Sequence Number15346701
Product Code DQO
UDI-Device Identifier00183739000654
UDI-Public00183739000654
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K141769 
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/17/2024
Device Catalogue NumberC408646
Device Lot Number8698124
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/02/2023
Supplement Date Received by Manufacturer11/15/2023
Initial Report FDA Received Date10/27/2023
Supplement Report FDA Received Date12/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-