| Model Number |
VELA CF PLUS,DISS,GRN O2,TYPE K,ENG OVERLAY ACCOUNT NAME |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.However, customer requested an fse (field service engineer) repair.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3: other text: device not returned yet.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that the vela ventilator had a circuit disconnect alarm back on (b)(6) while on a patient use.Furthermore, there was no patient harm reported associated with this event.Customer confirmed that the unit was replaced by a different ventilator.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: g3, g6, h2, h3, h6 and h11.Results of investigation: vyaire medical was able to verify the reported issue.The suspect device was not returned for investigation.However, as per work order performed under wo-00179310, it has been noticed that customer is using ex flow sensor for an avea unit with the longer tubes.The flow sensor receptacle is also incorrect.The flow sensor receptacle installed does not go with the unit.The diaphragm had the nipple facing inwards.Field service engineer corrected it all.Removed main board and installed new flow sensor receptacle for the vela unit.Installed new flow sensor.Reconnected main board and reassembled unit.Ran verification.All tests passed required criteria.Unit meets factory specifications.
|
| |
|
Search Alerts/Recalls
|