Model Number EUXHCP
Medical Device Problem Codes Increase in Pressure (1491); Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3 other text : device not returned to manufacturer.
 
Event or Problem Description
The manufacturer received information alleging a the device set pressure is 6 cm but when the device was calibrated the manometer showed a pressure of 10 cm and air was found to be leaking out of the unit.The device's humidifier plate was reported to be too hot to the touch and overheating.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The manufacturer received information alleging a the device set pressure is 6 cm but when the device was calibrated the manometer showed a pressure of 10 cm and air was found to be leaking out of the unit.The device's humidifier plate was reported to be too hot to the touch and overheating.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.**udi related data quality updates only** the udi in section d4 was previously reported in the incorrect format.The udi information has been updated to the correct format.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION HUMID
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18026353
Report Number2518422-2023-28043
Device Sequence Number7684567
Product Code BZD
UDI-Device Identifier00606959023246
UDI-Public(01)00606959023246
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberEUXHCP
Device Catalogue NumberEUXHCP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2023
Supplement Date Received by Manufacturer09/26/2023
Initial Report FDA Received Date10/27/2023
Supplement Report FDA Received Date08/21/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/29/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
-
-