|
This report is being supplemented to provide additional information, based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing, by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The event can be prevented, by following the instructions for use sections below: chapter 4: reprocessing workflow for endoscopes and accessories.Chapter 5: reprocessing the endoscope.In addition, the following non-reportable malfunctions were found, during the device evaluation: connecting tube dent, bending section wear and damage.Olympus will continue to monitor field performance for this device.
|