| Catalog Number |
03666794001 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The meter was requested for investigation and a new meter was sent to the customer as a replacement.Section e3 - occupation: patient/consumer.
|
| |
|
Event or Problem Description
|
|
We received an allegation of a display issue with a coaguchek meter.The reporter stated that the meter memory and segments were faded.The reporter replaced the batteries and stated that he could barely see the numbers in the top right.During the call, a display check was performed and the reporter stated that at first he could see everything but the segments and three 8's in the middle began to fade.The meter memory was also checked.The reporter stated that he can see the "mem" on the left and the "0" on the right but the three dashes in the middle were faded.The reporter stated that there was no corrosion in the battery compartment.
|
| |
|
Additional Manufacturer Narrative
|
|
Medwatch fields d9 device available for evaluation, d9 date device returned and h3 device not returned to mfg for eval updated.The meter was received for investigation.The investigation performed a display test; individual segments were shown faintly on the display.The investigation tested the printed circuit board (pcb) for damage and contamination; the battery contacts and pcb were contaminated by a leaked battery.The investigation determined the event was caused by contamination of the contacts due to improper handling or maintenance.Product labeling states: "- if you store the meter for a period of time, remove the batteries.-use the meter at a relative humidity of less than 85%, without condensation." "please follow the procedures below to clean and disinfect the meter.Failure to follow these procedures may cause malfunction of the meter.¿ do not use sprays of any sort.¿ ensure that swab or cloth is only damp, not wet." "wipe away residual moisture and fluids after cleaning the housing.".
|
| |
|
Search Alerts/Recalls
|