• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS GLUC3; HEXOKINASE, GLUCOSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS GLUC3; HEXOKINASE, GLUCOSE Back to Search Results
Catalog Number 05168791190
Medical Device Problem Codes High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The c 702 module serial number was (b)(6).The competitor urine strip was the sysmex uc-3500.
 
Event or Problem Description
The initial reporter questioned a high gluc3 assay result for 1 patient's urine sample tested on a cobas 8000 c 702 module.The urine results were >abs "several" times.When diluted x10, the urine result was 243.4 mmol/l.The result from a competitor urine strip was 111 mmol/l.The glucose result from the serum sample was 6.1 mmol/l.The customer is wondering about possible interferences that could affect the urine glucose result as it is not consistent with the serum glucose result.
 
Additional Manufacturer Narrative
The root cause of the event was found to be consistent with the actions of the drug jardiance.According to the drug's manufacturer, the drug blocks a specific protein in the kidneys which increases glucose secretion in urine.Based on the available data, a general reagent issue could be excluded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLUC3
Common Device Name
HEXOKINASE, GLUCOSE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18028330
Report Number1823260-2023-03458
Device Sequence Number20573374
Product Code CFR
UDI-Device Identifier04015630925025
UDI-Public04015630925025
Combination Product (Y/N)Y
Initial Reporter CountrySI
PMA/510(K) Number
K061048
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05168791190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/11/2023
Supplement Date Received by Manufacturer11/06/2023
Initial Report FDA Received Date10/30/2023
Supplement Report FDA Received Date11/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
JARDIANCE; R-CHOP; ZARZIO
Patient SexUnknown
-
-