| Model Number |
SONATA |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The child was brought by the parents for regular follow-up with the audiologist.Parents reported that the child responded to most of the sounds and common words (also observed during the session), but did not talk apart from vocalizing /aa/.No other technical checks have been scheduled.Diagnostic image report is awaited.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However, to determine an exact root cause a device investigation of the explanted device is necessary.No information on possible further steps has been received.
|
| |
|
Event or Problem Description
|
|
The child was brought by the parents for regular follow-up with the audiologist.Parents reported that the child responded to most of the sounds and common words (also observed during the session), but did not talk apart from vocalizing /aa/.No other technical checks have been scheduled.Diagnostic image report is awaited.
|
| |
|