ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1804350-38
Medical Device Problem Codes Break (1069); Leak/Splash (1354); Difficult to Remove (1528); Improper or Incorrect Procedure or Method (2017); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/06/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6: medical device problem code 2017- incorrect prep manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that the procedure was to treat the mid left anterior descending (lad) coronary artery with mild calcification, no tortuosity and 80% stenosis.The 3.5x38 mm xience skypoint stent delivery system (sds) was advanced to the lesion.When the tech pulled negative on the device and then tried to inflate the device, contrast began to spray out of a crack in the hub of the device.Negative pressure was pulled again and the sds was attempted to be removed; however, the stent dislodged from the balloon.A 2.0 mm balloon was then advanced and used to try and pull the stent back into the guiding catheter.The stent became caught on the moderate proximal disease in the vessel.The balloon was inflated in the distal part of the stent and since the stent was now partially deployed and there was disease in the proximal vessel, the balloon was used to deploy the stent where it was located and post-dilated with a 4.5 mm balloon to match the intravascular ultrasound (ivus) sizing of the vessel.The stent is located completely outside the target lesion; however, the area is diseased.A new 38 mm stent was implanted distally to the original lesion and was connected to the proximal stent.There were no reported adverse patient sequela.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported stent dislodgement, hub break and hub leak were confirmed.The reported difficulty to remove could not be replicated in a testing environment as it was related to operational context of the procedure.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties and treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.H6 - medical device problem code 2017 removed.
 
Event or Problem Description
It was reported that the procedure was to treat the mid left anterior descending (lad) coronary artery with mild calcification, no tortuosity and 80% stenosis.The 3.5x38 mm xience skypoint stent delivery system (sds) was advanced to the lesion.When the tech pulled negative on the device and then tried to inflate the device, contrast began to spray out of a crack in the hub of the device.Negative pressure was pulled again and the sds was attempted to be removed; however, the stent dislodged from the balloon.A 2.0 mm balloon was then advanced and used to try and pull the stent back into the guiding catheter.The stent became caught on the moderate proximal disease in the vessel.The balloon was inflated in the distal part of the stent and since the stent was now partially deployed and there was disease in the proximal vessel, the balloon was used to deploy the stent where it was located and post-dilated with a 4.5 mm balloon to match the intravascular ultrasound (ivus) sizing of the vessel.The stent is located completely outside the target lesion; however, the area is diseased.A new 38 mm stent was implanted distally to the original lesion and was connected to the proximal stent.There were no reported adverse patient sequela.Subsequent to the initially filed report it was reported that the contrast mix was 50/50 and the device was prepped outside the anatomy prior to use.When 2 atmospheres was reached during inflation, that is when the device began leaking at the hub.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18029867
Report Number2024168-2023-11948
Device Sequence Number6962294
Product Code NIQ
UDI-Device Identifier08717648233449
UDI-Public08717648233449
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/08/2026
Device Catalogue Number1804350-38
Device Lot Number3061941
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/06/2023
Supplement Date Received by Manufacturer11/14/2023
Initial Report FDA Received Date10/30/2023
Supplement Report FDA Received Date11/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;