Catalog Number B1080-060
Medical Device Problem Codes Material Rupture (1546); Material Separation (1562); Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.H6: device code 2017 clarifier - above rated burst pressurena.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the hepatic trunk artery.The 8x60mm armada 35 percutaneous transluminal angioplasty (pta) catheter met resistance during advancement due to the lesion; however, was inflated twice up to 12 atmospheres and a rupture occurred.During removal the device separated into two pieces and a fragment became stuck on the wire; therefore, a loop snared was used to retrieve the separated balloon fragment.There were no reported adverse patient sequela and no clinically significant delay.No additional information was provided.
 
Additional Manufacturer Narrative
Visual inspection was performed on the returned device.The reported balloon rupture and separation were confirmed.The reported difficulty to advance could not be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulty advancing the device, separation and unexpected medical intervention appear to be related to operational context; however, a conclusive cause for the reported balloon rupture could not be determined.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.H6- device code 2017 was removed.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the hepatic trunk artery.The 8x60mm armada 35 percutaneous transluminal angioplasty (pta) catheter met resistance during advancement due to the lesion; however, was inflated twice up to 12 atmospheres and a rupture occurred.During removal the device separated into two pieces and a fragment became stuck on the wire; therefore, a loop snared was used to retrieve the separated balloon fragment.There were no reported adverse patient sequela and no clinically significant delay.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18030707
Report Number2024168-2023-11959
Device Sequence Number5954623
Product Code LIT
UDI-Device Identifier08717648154874
UDI-Public08717648154874
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Catalogue NumberB1080-060
Device Lot Number30222G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/06/2023
Supplement Date Received by Manufacturer12/19/2023
Initial Report FDA Received Date10/30/2023
Supplement Report FDA Received Date01/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Patient Weight130 KG
-
-