• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS 400 MICRON FIBER PROCEDURE KIT; VENACURE ENDOVENOUS LASER TREATMENT FIBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS 400 MICRON FIBER PROCEDURE KIT; VENACURE ENDOVENOUS LASER TREATMENT FIBER Back to Search Results
Catalog Number EVLTPVAK
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/09/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.Although the samples were not returned the customer provided a photo of the fiber noted to have fractured and detached.The customer's reported complaint description of the fiber was fractured and detached was confirmed via picture provided by the customer, however, no sample was returned.Without receiving a sample for evaluation the root cause could not be determined.Potential root cause could be due to handling damage during transit and/or storage at the customer facility.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Manufacturing personnel 100% visually inspect devices during the packaging process.This type of fiber kink/detached damage would be noticed prior to shipment.Labeling review: the instructions for use which is supplied to the end user with this catalog number, states: "prior to and during use, avoid damaging the fiber by striking, stressing, or excessive bending.Do not coil the fiber tighter than a diameter of 20cm.Clinical safety and effectiveness data is not available for other fiber tip designs and diameters.Prior to and during use, avoid bending the introducer sheath and dilator as this can cause kinks and damage".A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Event or Problem Description
An end user reported an issue with a pvak -- 400 micron perforator and accessory vein ablation kit.While unpacking the device, the fiber was found to be fractured into two separate pieces.There was no damage noted to the internal or external packaging of the device.The following procedure was completed with another same device and the patient did not experience any adverse effects, harm, or require medical intervention as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
400 MICRON FIBER PROCEDURE KIT
Common Device Name
VENACURE ENDOVENOUS LASER TREATMENT FIBER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key18032184
Report Number1319211-2023-00072
Device Sequence Number6074108
Product Code GEX
UDI-Device IdentifierH787EVLTPVAKUS5
UDI-PublicH787EVLTPVAKUS5
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K041957
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Catalogue NumberEVLTPVAK
Device Lot Number5775577
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2023
Initial Report FDA Received Date10/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-