• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. STFX SYM PDS+ UNI VIO 24IN 0 S/A CT; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. STFX SYM PDS+ UNI VIO 24IN 0 S/A CT; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE Back to Search Results
Catalog Number SXPP1A441
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a thyroidectomy procedure (b)(6) 2023 and suture was used.Before use on the patient, it was reported that the problem of pulling off suture needle had happened in the newly dispensed suture when it was opened to prepare for surgery.Changed another one to complete the surgery.The product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that it was received one needle and suture that pertain to product code sxpp1a441.After investigation of the sample, short insertion was observed in the strand.The visual inspection concluded that the combination of the needle/suture was detached from the needle since the insertion end of the suture did not meet the requirement.Based on the information currently available, short insertion issue was identified as pull out during the investigation of the sample received.No adverse patient consequences were reported.No additional information was provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigation summary:the product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that it was received one needle and suture that pertain to product code sxpp1a441.After investigation of the sample, short insertion was observed in the strand.The visual inspection concluded that the combination of the needle/suture was detached from the needle since the insertion end of the suture did not meet the requirement.Based on the information currently available, short insertion issue was identified as pull out during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STFX SYM PDS+ UNI VIO 24IN 0 S/A CT
Common Device Name
SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key18035595
Report Number2210968-2023-08353
Device Sequence Number8398189
Product Code NEW
UDI-Device Identifier10705031244962
UDI-Public10705031244962
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K141776
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSXPP1A441
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2023
Initial Report FDA Received Date10/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-