• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO COLONOSCOPE 4.2C 13.2T 1700L FWJ

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO COLONOSCOPE 4.2C 13.2T 1700L FWJ Back to Search Results
Model Number EC-3890LK
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the suction nipple loose.Based on the result, we concluded that it was caused due to the excessive force applied on the suction nipple.In addition, our technician confirmed that the suction nipple leak, the remote control buttons cut, the ground wire cut, the ccd module with drive pcb pixels, and the air/water socket chipped/cut o-ring; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0588(channel)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Suction nipple loosened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
VIDEO COLONOSCOPE 4.2C 13.2T 1700L FWJ
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315800
MDR Report Key18036671
Report Number9610877-2023-60350
Device Sequence Number8166920
Product Code FDF
UDI-Device Identifier04961333129447
UDI-Public04961333129447
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K131855
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEC-3890LK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2023
Initial Report FDA Received Date10/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-