Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
It was reported that caller had seen the patient yesterday for normal check in and noted the impedances were a little elevated on the left side 0/2 5337, 0/3 5471m c/0 5139 but 0/1 were normal and that is what they were programmed on.Impedances from 6 months ago were normal.Reviewed considerations and since the patient is getting great therapy they will be keeping an eye on it.Patient is at 2.68v battery level so when they go to replacement she will suggest checking the components.No symptoms were reported.Additional information received from the healthcare provider (hcp) reported they were with the patient today and a impedance check at 3 volts showed investigate for electrode 0 with results of >4,000 ohms.Patient was in group b with 0+ and -1 at 90usec.It was reviewed that given the impedance was high they may not want to use electrode 0 and to try c1 or 1- and 2+ coupled with pulse width and rate adjustment to see if the patient would get relief.The hcp decided to make a new group.The cause of the high impedance was unknown and not resolved.
|