Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).H3: the complaint device has been returned, but the device investigation has not yet been completed.Once the evaluation is completed, a supplemental medwatch will be submitted.
 
Event or Problem Description
It was reported that the device was discovered to have missing components.Attempts to obtain additional information have been made; however, no more is available.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual examination of the returned product identified signs of use and shim is missing ball bearing.Reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.Complaint is confirmed based on the product evaluation.If any further information is received which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM
Common Device Name
KNEE INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18038052
Report Number0001822565-2023-03022
Device Sequence Number12831511
Product Code JWH
UDI-Device Identifier00889024245778
UDI-Public(01)00889024245778(10)63900971
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42527900303
Device Lot Number63900971
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2023
Supplement Date Received by Manufacturer12/07/2023
Initial Report FDA Received Date10/31/2023
Supplement Report FDA Received Date12/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-