| Catalog Number |
19AGFN-756 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/22/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2023, 19mm sjm regent heart valve was implanted in a patient.At the time of device preparation, the leaflets was moving normally when tested.After 3 days of the implantation, the doctor found out the mechanical lobe did not work well and may get stuck.The doctor did a secondary thoracotomy operation on (b)(6) 2023.Device was replaced with a non-abbott device.Post explant the device was moving normally when tested.The patient is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of the valve abnormally coapting and explant was reported.The device was received for examination and no anomalies were found.The valve was sent for functional testing which confirmed that the valve functioned normally, with the leaflets moving with no asynchronous motion.It was indicated that post explant the device was moving normally when tested.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|