Catalog Number 04404483190
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The cobas 6000 c 501 serial number was (b)(6).The investigation is ongoing.
 
Event or Problem Description
There was an allegation of a questionable gluc hk gen.3 (glucose) result from the cobas 6000 c 501.The initial result was 2 mg/dl and was reported outside of the laboratory.The repeat result was 77 mg/dl and was believed correct.
 
Additional Manufacturer Narrative
The customer canceled a scheduled service visit and stated they resolved the issue by recalibrating the assay.The investigation determined the customer's actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS INTEGRA GLUCOSE HK GEN. 3
Common Device Name
HEXOKINASE, GLUCOSE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18039245
Report Number1823260-2023-03475
Device Sequence Number20932241
Product Code CFR
UDI-Device Identifier04015630920297
UDI-Public04015630920297
Combination Product (Y/N)Y
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K061048
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04404483190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/12/2023
Supplement Date Received by Manufacturer11/16/2023
Initial Report FDA Received Date10/31/2023
Supplement Report FDA Received Date11/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-