| Catalog Number |
04404483190 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The cobas 6000 c 501 serial number was (b)(6).The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of a questionable gluc hk gen.3 (glucose) result from the cobas 6000 c 501.The initial result was 2 mg/dl and was reported outside of the laboratory.The repeat result was 77 mg/dl and was believed correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer canceled a scheduled service visit and stated they resolved the issue by recalibrating the assay.The investigation determined the customer's actions resolved the issue.
|
| |
|