| Catalog Number |
962062 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Synovitis (2094); Joint Laxity (4526); Insufficient Information (4580)
|
| Date of Event |
09/20/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Clinical adverse event received for poly failure, poly wear.Event is serious and is considered moderate.Event is possibly related to device.Event is not related to procedure.Date of implant: (b)(6) 2008.Date of revision: (b)(6) 2023.Date of event: (b)(6) 2023.(left knee).Treatment: revision; insert was revised.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).The mfg date is unknown at this time.If follow-up information is received, a follow-up regulatory report will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Operative notes reviewed.On (b)(6) 2023 the patient had a revision left total knee replacement to address wear of the articular bearing surface, internal derangement of left knee.It was noted that the patient had ligamentous laxity and there was a minimal amount of foreign body synovitis.The patella was noted to have small amount of wear, but was found to be satisfactory and retained.Only the insert was revised.The operative findings included polyethylene wear tibial component and patella component.Depuy insert was implanted during this procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H10 additional narrative:.
|
| |
|
Search Alerts/Recalls
|